Skip to content
Talk to usNeed help?Get Started

Transform sensor data

Apply a depth-to-water calibration

Apply a depth-to-water template to a soft sensor using matching reference measurements.

A DTW calibration converts level readings into a depth to water measurement. It is intended for users with super user access who are responsible for configuring sensor data transformations.

Make sure the following is true before you begin:

  • You have super user access. This is required to create soft sensors and calibrations.
  • You have already created the soft sensor that this calibration will be applied to. If you have not done this yet, see Creating a soft sensor in Orion Data. Soft sensors are created from the Channels tab.
  • You have your reference measurements ready: the Dip and the Head of water at time of dip.
  1. Go to Devices and select the device that contains your soft sensor.
  2. Click the Calibration tab.
  3. From the Select Sensor dropdown, choose the soft sensor you created for the DTW measurement.

Select Sensor dropdown open on the Calibration tab, with the Water Level soft sensor highlighted

  1. Click Create From Template.
  2. In the search field, type DTW.
  3. Select the DTW template from the results.

Create From Template dropdown searched for DTW, with a DTW(m) Calc template highlighted

The DTW template requires two reference values taken at the same point in time:

  • Dip - the manually measured depth to water at the time of the reference reading.
  • Head of water at time of dip - the corresponding sensor level reading captured at the same moment as the dip.
  1. Enter the Dip value.
  2. Enter the Head of water at time of dip value.
  3. Click Save.

DTW calibration steps with the Dip and Head of water at time of dip value fields highlightedDTW calibration steps with an arrow to the Save button

Saving the calibration applies it to new readings going forward. To apply the calibration to data already recorded against the sensor, run a force-recalculation.

For instructions, see Using Force-recalculation.

Calibration Step Groups with the DTW(m) Calc group and the Force-recalculation button highlighted

This article covered how to apply a DTW calibration to an existing soft sensor in Orion Data. You opened the Calibration tab for the device, selected the soft sensor, applied the DTW template using Create From Template, entered the Dip and Head of water at time of dip values, and saved. Running a force-recalculation then applies the calibration to previously recorded data.

  • Verify the calibrated output by viewing the soft sensor’s recent readings.
  • Run a force-recalculation to apply the calibration to historical data. See Using Force-recalculation.
  • Configure alarms on the soft sensor if you need to be alerted to defined depth to water conditions.